Metrohm IC Net 2.0 User Manual

Page 40

Advertising
background image

5 Interfaces

IC Net 2.0

32

5.2.4

762 IC Interface window

762 icon / Open

The

762 IC Interface

window for parameter settings is opened by

selecting this menu option with the right mouse button or by

clicking the

762 icon

on the toolbar. It consists of three tabs

Init

,

RS settings

, and

Links

.

Init

The

Init

tab of the

762 IC Interface

window contains data acquisi-

tion parameters for the 762 IC Interface.

Label

Optional label to name the interface with

maximum 8 characters.

ROM version

Instrument program version number.

Measuring per second

Number of data points measured per s.

Entry range:

10, 20, 30, 50, 60 points/s

Range

Range for AD converter.
Example: converts

±

2500 mV to 2

24

bits

Entry range:

2500, 1250, 625, 312.5, 156.25,

78.125, 39.062 mV

Second press on RUN/STOP means 'stop' when measuring

If this option is enabled, the running data

acquisition will be stopped if the

RUN/STOP button on the 762 IC Interface

front panel is pressed. This is equal to

Stop

determination

.

If this option is disabled, the running data

acquisition will not be stopped if the

RUN/STOP button on the 762 IC Interface

front panel is pressed.