Metrohm IC Net 2.0 User Manual

Page 53

Advertising
background image

5.4 761 Compact IC

IC Net 2.0

45

Second column

Program instruction (see below).
In addition to these predefined instructions,

user-defined remote commands can be

entered if activated on the

Remote configura-

tion

tab.

Third column

Parameter for program instruction (see

below).

ENABLED

Enable program start (a disabled program

is not started).

<Add>

Add new program instruction.

<Delete>

Delete selected program instruction.

<Verify>

Test the time program (error messages are

displayed if program is wrong).

List of program instructions

The following program instructions can be added to the time pro-

gram on the

Program

subpage:

Instruction

Parameter entry

Meaning

Valve

Inject, Fill

Switch injection valve to "INJECT"

or "FILL" position.

FullScale

50, 250, 1000
µµS/cm

Set full scale range to the se-

lected value.

ICPump

on, off

Switch on or off the high-pressure

pump.

Flow

0.2 ... 2.5 mL/min

Set flow rate of the high-pressure

pump to the desired value.

Pmax

0.0 ... 25.0 MPa

Set maximum pressure limit for

the high-pressure pump to the de-

sired value.

Pmin

0.0 ... 25.0 MPa

Set minimum pressure limit for

the high-pressure pump to the de-

sired value.

Remote

0, 1, *, p

Set remote output lines 1...8 to

the desired values. For entry of the

first value, enter

1

,

0

,

p

or

*

. For en-

try of the other values, move the

cursor in front of the value to be

changed and enter

1

,

0

,

p

or

*

.