Metrohm IC Net 2.0 User Manual

Page 57

Advertising
background image

5.4 761 Compact IC

IC Net 2.0

49

Cell constant

Cell constant of the conductivity cell for

correct display of the absolute conductivity.

Enter the cell constant printed on the de-

tector block into this field.

Range:

0.1

...

1000 /cm

For a precise determination of the cell con-

stant, pump a calibration solution of known

conductivity through the IC system, ob-

serve the displayed conductivity and

change the cell constant until the correct

conductivity value is displayed.

<Break times>

Possibility for changing the break times for

injection valve

Valve

and suppressor mod-

ule

Suppressor

.

These values should only be changed af-
ter consulting the Metrohm Service.

Outputs

The

Outputs

tab of the hardware settings window defines remote

output signals to be set automatically if specific events occur.

Event

Events for automatic remote signal output:

Inject

Switching of the injection valve to the

"INJECT" position.

Fill

Switching of the injection valve to the "FILL"

position.

Suppressor step

Switching of the suppressor module to the

next position.

Pump stop (P max/min)

Pump stopped because pressure limits are

exceeded.