background image

7 Methods

IC Net 2.0

116

Integration events for peak start/end

Set peak start

Force the beginning of a new peak. If a

peak is available at the moment of the

event it is either rejected (upslope) or ter-

minated (downslope).

Set peak end

Force peak end at the time of the event.

Upslope peaks are rejected (except those

born by

Set peak start

event), downslope

peaks are terminated.

Stop single peak mode

Disable peak end detection. All peaks of

any group after this event will be treated as

one peak. The group end will be treated as

peak end.

Start single peak mode

Set normal detection mode, when valley

causes perpendicular drop or skim line

separation.

Split peak

Terminates the current peak and starts a

new one.

Integration events for baseline

Enable valley-to-valley Disable valley-to-valley

Enable valley-to-valley

Disable perpendicular drop peak separa-

tion. All peaks are considered to be base-

line-separated. The bottom of the valley

becomes the baseline point.

Disable valley-to-valley

Enable perpendicular drop peak separation

(default setting).

Set horizontal Set back hori-

Set normal

baseline

zontal base

baseline