background image

7 Methods

IC Net 2.0

130

Coef. (balance)

Recalculate coefficient of the universal com-

ponent. The aim of this action is to adjust the

response factor of this component so, that the

total amount of the injected components be-

comes equal to the declared value. This de-

clared summary amount is entered into the

Concentration

field of the universal component

for

This run

level.

7.5.9

Calibration data handling

Load and save calibration data

IC NET / Method / Calibration / Load from method

Load the calibration from the method. This option is used to up-

date the calibration for the current run by the one taken from the

method used for acquisition.

IC NET / Method / Calibration / Save to method

Save the current calibration to the method. This option is used to

transfer the updated calibration from the run to the current

method (i.e. the method associated with the current chroma-

togram window).

Import and export calibration data

IC NET / Method / Calibration / Import calibration

This option imports a calibration from a calibration file

*.cal

to the

current method or chromatogram. It is used to transfer the up-

dated calibration from chromatogram-to-chromatogram or from

method-to-method. Different methods can be processed in this

case using the same calibration.

IC NET / Method / Calibration / Export calibration

This option exports a calibration to a in calibration file

*.cal

. It is

useful to transfer a calibration from chromatogram-to-

chromatogram that were obtained by different methods.

Put out calibration curves

COMPONENT / Copy to clipboard

This option allows to copy the calibration curve of the selected

component to the clipboard so that it is available for other Win-

dows applications, such as WinWord, Excel, etc.

COMPONENT / Print/Preview / Preview this

Display calibration curve print preview of the selected compo-

nent.

COMPONENT / Print/Preview / Preview all

Display calibration curve print preview of all components.

COMPONENT / Print/Preview / Print this

Print selected calibration curve of the selected component.

COMPONENT / Print/Preview / Print all

Print calibration curves of all components.