7 Methods

IC Net 2.0

136

Channel table

Parameters of the data acquisition channel

used.

CHANNELS TABLE
ch1
No Name Units Input Minimum Zero Maximum Range Coefficient Noise Shift

1 ch1 mV 1 -8388607 0 8388607 2.500e+03 2.980e-04 4.15 0

# Name/Units Noise RMS PeakToPeak Drift/hour
1 ch1 4.2 29976.5 330509.6 3585.801
mV 0.00124 8.93 98.5 1.07

Name

Channel’s name. This text appears as

Channel label

in the chromatogram.

Units

User-defined units of detector response.

Input

Number of data acquisition channel.

Minimum

Minimum value of the linear range of the

AD converter (in ADC conversion units).

This parameter is used to detect an under-

flow condition.

Zero

A signal level on a baseline (in ADC con-

version units). This parameter is a result of

the ADC calibration.

Maximum

Maximum value of the linear range of the

AD interface (in ADC conversion units). This

parameter is used to detect overflow con-

dition.

Range

An input signal value in maximum in user-

defined units.

Range

= (

Maximum

-

Zero

)

Coef

Coefficient

ADC sensitivity coefficient (weight of ADC

bit in user-defined units).

Noise

Estimated baseline noise value of the

channel in ADC conversion units (bits).

Shift

Shift of the current channel in time. This

parameter is useful in the case of multi-

channel chromatograms only.

(multi)Channel ratio

Spectral ratios. The option is valid for multi-

channel chromatograms only.