Metrohm IC Net 2.0 User Manual

Page 130

Advertising
background image

7 Methods

IC Net 2.0

122

Number

Row number (read-only).

Peak

Number of the peak in the chromatogram

that corresponds to the given component.

If this field contains a

0

, the peak number

must be entered manually.

Time

Expected retention time of the compo-

nent in the calibration run. After a new entry

of the peak number the corresponding re-

tention time is entered automatically as

soon as this field is clicked on with the

mouse.

Wind.%

Identification window of the component.

This value is the maximum allowed differ-

ence of the actual retention time of the

component and its expected retention time,

measured as % of expected retention time.

The component will be identified within its

identification window only.

Ref.

Reference component.

The entry

Yes

de-

fines components as reference compo-

nents; these ensure improved peak identifi-

cation of the other

Ordinary components

(details see on-line help).

Name

Component’s name (cannot include

spaces).

Group

Number of the group to which the compo-
nents are allocated. If group numbers

1

are entered here then a group report can

be outputted for each group; this also

contains the intermediate totals for this

group. This group report is produced after

the main report table.

Index

Retention index for components with

known index. In the case of an unknown in-

dex this value should be equal to

0

. For in-

dex calculation the user must define index

values for at least two peaks, and all other

values will be calculated by the software

using linear or logarithmic approximation

(details see on-line help).

RF

Response factor.

This value corresponds

to the slope of the calibration curve

(coefficient k

1

of the calibration formula).

Advertising