Metrohm IC Net 2.0 User Manual

Page 56

Advertising
background image

5 Interfaces

IC Net 2.0

48

5.4.5

Hardware settings

Selection of the

Hardware

menu item of the

761 icon

using the right

mouse button opens the hardware settings window consisting of

the three tabs

Hardware

,

Outputs

and

Alarm stops

.

Hardware

The

Hardware

tab of the hardware settings window defines gen-

eral parameters which are set automatically at power on of the

761 Compact IC.

Remote lines after power on

The remote output lines 1

8 are set to this

values after power on or a manual stop with

Shutdown hardware

.

Selection:

0

,

1

Thermostat

Operating temperature of the conductivity

cell.

Selection:

25

,

30

,

35

,

40

,

45 °C

,

off

Thermostatting functions only if the ambient tem-
perature is at least 5 °C lower than the operating tem-
perature. It normally takes 30

…60 min after power on

until a temperature stability of

± 0.01 °C is attained.

Flow correction

Factor for correction of the difference be-

tween displayed and actual flow rate of the

high-pressure pump.

Range:

0.9

...

1.09

The flow correction is determined by meas-

urement of the actual flow rate with the aid

of a measuring cylinder as follows:

Flow correction

=

Displayed flow rate
Measured flow rate