5 Interfaces

IC Net 2.0

46

Program

END, RESET

The

END

flag can be used to end a

program, especially if the program

time should be longer than the

chromatogram duration. Additional

steps after this flag are not allowed.

The

RESET

flag is used to reset the

parameters to the system startup

values.

Suppressor

Switch suppressor module to the

next position.

Peristaltic

on, off

Switch on or off the peristaltic

pump.

Remote configuration

On the

Remote configuration

subtab user-defined remote com-

mands can be defined, which can be inserted into a time pro-

gram after being activated with

<Activate>

.

Name (1

st

column)

User-definable name of the remote com-

mand (e.g.

Start_766

).

Remote command (2

nd

column)

Setting the remote output lines 1

8.

0

(line off, inactive, open)

1

(line on, active, 0 V)

p

(pulse)

*

(leave line in current status)

For entry of the first value, enter

1

,

0

,

p

or

*

.

For entry of the other values, move the cur-

sor in front of the value to be changed and

enter

1

,

0

,

p

or

*

.

<Add>

Add new remote command.

<Delete>

Delete selected remote command.

<Activate>

Activate the defined remote commands for

insertion into the time program.