Appendix A – Configuring a 1-5Vdc input for a 0-250 psig pressure transducer

Page 3 of 6

Using the up/down arrow keys select

Output scale

and

touch

ENTER

.

You will now be prompted for the number of decimal places

that you would like to display and record. Using the up/

down arrow keys you can select 0, 1, 2, 3, 4, or 5decimal

places. For this example select 2 and touch

ENTER

.

You will now be prompted for the

hi

end of your output

scale. You should enter

250.00

and touch

ENTER

.

Another numeric keypad will appear prompting you for the

lo

end of your output scale. For this example (0-250 psig)

enter

0.00

and touch

ENTER

.

11.

12.

13.

14.

The numeric keypad will now prompt you for the

hi

end of

your input signal. For this example enter

5.00

and touch

ENTER

.