2 installation – Monarch Instrument DC 2000 User Manual

Page 20

Advertising
background image

Page 2-2

Chapter 2 Installation and Wiring

2.2 Installation

The instrument is intended to operate in the following environment:

Indoor Use Only
Installation Category II

per IEC 664

Pollution Degree Level II per UL3111-1/IEC1010-1
Temperature

5° C to 40° C per UL3111-1/IEC1010-1 with disk drive
-10° C to 50° C with PCMCIA drive or Zip drive

Humidity

10% to 80% RH per UL3111-1/IEC1010-1

AC Mains Supply

100 - 240 Vac ~ 50/60 Hz 35 Vac

NOTE:

The recorder is designed to be panel mounted and as such should be considered as permanently
connected. Disconnection from the supply must be possible via a customer supplied switch or circuit
breaker. This disconnection device must be included in the panel installation and should be clearly
marked, in close proximity to the Recorder and easily accessible to the operator.

The Recorder is sized to fit in a standard DIN panel cutout of 5.43” x 5.43” [138 mm x 138 mm] and requires 7.41”
[188.2 mm] behind panel depth not including power and input source cable space needed. Actual dimensions are
shown in Figure 2-1.

Figure 2-1 Recorder Dimensions in Inches [mm]

WARNING: Do not use this instrument in any manner inconsistent with these operating

instructions or under any conditions that exceed the environmental
specifications stated.

Advertising