AppendixA – Configuring a 1-5 Vdc input for a 0-250 psig pressure transducer

Page 5 of 6

The scale setup menu will be displayed. Use the up/down

arrow keys to select

Scale ends

and touch

ENTER

.

You will now be prompted with

places?

(0-5 will also be

displayed). Use the up/down arrow keys to select 2 and

touch

ENTER.

This selects the number of decimal places

that will be displayed in trend view only.

You will now be prompted with

mid

(some #). Here you

can enter the mid point of your scales range. For this

example (0-250 psig) enter

125.00

and touch

ENTER

.

You will now be prompted with

lo

(some #)

.

Enter the low

end of your scale. For this example (0-250 psig) enter

0.00

and touch

ENTER

.

21.

22.

23.

24.

You will be prompted with

scale?

(1A-2H will also be

displayed) Use the up/down arrow keys to scroll through the

different scales until

1A

is displayed and then touch

ENTER

.