Appendix C – Configuring a J type T/C input for a 100 - 750°C range

Page 4 of 5

You will now be in the scale menu. Use the up/down arrow

keys to select

Scale ends

and touch

ENTER

. You will be

prompted with

places

? (for decimal places). Using the up/

down arrow keys you can select 0, 1, 2, 3, 4 or 5. For

thermocouples you would only select 0 or 1. For this

example select

1

and touch

ENTER

.

You will be prompted with

lo

→ ←

. For this example

(100 to 750 °C) enter

100

for the low range of your scale

and touch

ENTER

.

The next prompt will be

hi

→ ←

. For this example enter

750

and touch

ENTER

. You will now be back in the main

scale menu. Use the up/down arrow keys to select

Scale

units

and touch

ENTER.

You will now be prompted with

mid

→ ←

. To illustrate one

ofthe features ofthe Recorder we will enter a value that is

not 50% ofthe desired range. For this example, enter

200

and touch

ENTER

. The result will be two linear ranges.

One will have high resolution and the other will have low

resolution.

Note

: Ifyou wanted the scale to be linear you would simply

enter the exact middle ofyour range at this point.

16.

17.

18.

19.

Use the up/down arrow keys to select

Scales

and touch

ENTER

. You will be prompted

scale?

. Use the up/down

arrow keys to select

1A

and touch

ENTER.