Chapter 2 Installation and Wiring

2.4 Serial Interface Option

Data can be accessed to download Configuration files or Data files using the RS232C option and a modem. The
RS485 option allows the Recorder to be installed into an existing Modbus network or it can be used to connect up
to thirty-one recorders in series. A standard DB9 Female connector is required for the RS232 to connect to an IBM
PC compatible computer using a null modem cable and the RS232 can support cable runs up to 50 feet [16 m]. The
RS485 connection is via two wire (twisted pair) cable (a DB9 Female connector is required) and can support cable
runs up to 4000 feet [1300 m].

The Serial Interface contains an isolated switching unit for RS232 and RS485 access with a standard DB9 Female
connector. When switch 2 is in the ON position, RS485 is enabled. When switch 2 is in the OFF position, RS232
is enabled.

When more than one recorder are connected in a series, it is necessary to apply a termination resistor on the last
recorder. Switch 1 in the ON position applies this necessary termination resistor and should be switched to the ON
position only on the last recorder in series.

Figure 2-10 RS232/RS485 Modbus

All Serial Interface connections are made through the DB9 female connector.

The RS232 Connection to the DB9 female connector are as follows:

DB9 PIN

CONNECTION

DIRECTION

2

RxD Receive Data

In

3

TxD Transmit Data

Out

4

DTR Data Terminal Ready

Out

5

Common

N/A

7

RTS Request To Send

Out

8

CTS Clear to Send

In

The RS485 (Half Duplex Mode) Connection to the DB9 female connector are as follows:

DB9 PIN

CONNECTION

DIRECTION

6

A- Negative

Input/Output

9

B+ Positive

Input/Output

SW1

On=485 Term.

SW2

On=RS485
Off=RS232

1 2

ON