Appendix C – Configuring a J type T/C input for a 100 - 750°C range

Page 3 of 5

You will now be back at the menu starting with

Point tag

.

Use the up/down arrow keys to select

Compensation

and

touch

ENTER

. Use the up/down arrow keys to select

Local

comp

and touch

ENTER

. The Recorder will now use a

temperature sensor located at the rear of the recorder to

correct for cold junction error. You must touch

EXIT

to

leave this menu.

You should now be back at the

Point tag

menu. Use the up/

down arrow keys to select

Eng units

and touch

ENTER

.

You will be prompted with

units

(°C or °F). For this

example use the up/down arrow keys to select

°C

and touch

ENTER

.

Use the up/down arrow keys to select

Chart/Pens

and touch

ENTER

.

You will again be in the

Point tag

menu. Use the up/down

arrow keys to select

Chart scale

and touch

ENTER

. You

will be prompted with

chart scl=

. Use the up/down arrow

keys to select

A

and touch

ENTER

. Now touch

EXIT

. You

will be prompted with

keep setup?

. Touch

YES

and you

will be returned to the main programming menu.

11.

12.

13.

14.

The menu starting with

Point tag

will again be displayed.

Use the up/down arrow keys to select

Decimal fix

and touch

ENTER

. You will be prompted with

places>?

(0 or 1). Use

the up/down arrows to select zero or one decimal place for

thermocouples. For this example select

1

and touch

ENTER

.

10.