Activating the monitor – ADS Environmental Services FlowAlert QR 775011 A2 User Manual

Page 128

Advertising
background image

5-24

ADS FlowAlert Manual

Activating the Monitor

After configuring the monitor, activate the monitor to initiate the
flow monitoring process based on the monitor configuration.
Monitor activation involves generating the activation data using the
Profile software, downloading this data to the monitor, and
initiating flow data measurement and logging. The activation data
includes the current monitor firmware (embedded software) and
relevant portions of the LIF and other configuration parameters
necessary to ensure monitoring activities reflect the specific site
conditions and project requirements. The monitor requires these
files and information to properly measure and record the flow.

Activating the monitor also tests communication between the user's
computer and the monitor. To test communication, connect to the
monitor on-site (using the Direct Connection cable) or have an
analyst call the monitor from any telephone or computer at a remote
location while a technician is on-site to verify that the monitor
modem responds. This also will verify the monitor telephone
number entered in Profile.

Monitor activation occurs through the Diagnostics tool in Profile.
Activate the monitor in the following way:

1. Select the monitor location for activation from the Profile main

screen, and then click on the Diagnostics toolbar button.

The Diagnostics dialog displays.