AERCO RECON 1000 User Manual

Page 62

Advertising
background image

RECON Water Heaters Installation, Operation & Maintenance Manual

CHAPTER 4 – INITIAL START-UP

Page 62 of 178

AERCO International, Inc.

 100 Oritani Dr.  Blauvelt, NY 10913

OMM-0103_0A

PRI: 10/28/2014

Phone: 800-526-0288

GF-147

NATURAL GAS Combustion Calibration – Continued

NOTE

If NOx readings exceed the target values shown (<20 ppm), increase the O

2

level up to 25% higher than the listed calibration range shown in the table.
Record the increased O

2

value on the Combustion Calibration sheet.

Combustion Calibration Readings

Valve

Position

Oxygen (O

2

) %

Nitrogen

Oxide (NOx)

Carbon

Monoxide (CO)

Flame

µA

80%

6.5% ± 0.2%

<20 ppm

<100 ppm

> 6

60%

7.5% ± 0.2%

<20 ppm

<100 ppm

> 6

50%

8.0% ± 0.2%

<20 ppm

<100 ppm

> 6

40%

8.0% ± 0.2%

<20 ppm

<100 ppm

> 6

30%

8.3% ± 0.2%

<20 ppm

<100 ppm

> 5

20%

8.0% ± 0.2%

<20 ppm

<100 ppm

> 4

16%

5.8% ± 0.2%

<20 ppm

<100 ppm

> 4

23. If the oxygen level at the 16% valve position (RECON 500) or 20% valve position (RECON

1000) is too high and the Blower voltage is at the minimum value, you can adjust the idle
screw (TAC valve), which is recessed in the top of the Air/Fuel Valve (Figure 4-5). Rotate the
screw 1/2 turn clockwise (CW) to add fuel and reduce the O

2

to the specified level.

Recalibration MUST be performed again from 60% down to 16% after making a change to the
idle screw (TAC valve).

Figure 4-5. TAC Valve Adjust (Idle Screw)

This completes the Natural Gas combustion calibration procedure.

 
 

4.3.1 REASSEMBLY AFTER COMBUSTION CALIBRATION

Once the combustion calibration adjustments are properly set, the unit can be reassembled for
service operation.

TAC VALVE

IDLE SCREW

FRONT VIEW – POWER BOX REMOVED FOR CLARITY

Advertising
This manual is related to the following products: