Series 3L Indicators

11

3. Switch On

3.1 New Indicator

If the indicator is new and has not previously been
configured, or following a ‘Cold Start’ (section 5.5), it
will start up showing the ‘Quick Configuration’ codes.
This is a built in tool which enables you to configure
the input type and range, the output functions, and the
display format.

!

Incorrect configuration can result in damage

to the process and/or personal injury and must be
carried out by a competent person authorized to
do so. It is the responsibility of the person
commissioning the instrument to ensure the
configuration is correct.

The quick code consists of two ‘SETS’

of five characters.


The upper section of the display shows the set
selected. The lower section shows the five digits which
make up the set.
Adjust these as follows:

1. Press any button

.

The first character will change

to a flashing ‘-‘.

2. Press

or

to change the flashing character

to the required code shown in the quick code
tables –see next page. Note: An

x

indicates that

the option is not fitted.

3. Press

to scroll to the next character.

You cannot scroll to the next character until the

current character is configured.

To return to the first character press

.

4.

When all five characters have been configured the
display will change to

rng.hi followed by rng.lo

which allows range high and low limits to be set.

5.

The next press of

will select

Set 2. Adjust

each character as described for Set 1.

6.

When the last character has been entered press

again;

the display will show

. Continue

to press

if you wish to repeat the above quick

codes or press

or

to

if you are

satisfied with the quick codes. The indicator will
then automatically go to the operator level

SET 1

K C H C 0

Input Type

Thermocouple

B

Type B

J

Type J

K

Type K

L

Type L

N

Type N

R

Type R

S

Type S

T

Type T

C

Custom C

RTD

P

Pt100

Linear (all units)

M

0-80mV

2

0-20mA

4

4-20mA

Display
units

Decimal point

Temperature

0

nnnnn

(1)

X

None

1

nnnn.n

(1)

C

o

C

2

nnn.nn

(1)

F

o

F

3

nn.nnn

(1)

K

K

4

n.nnnn

(1)

P

%

Not Applicable

X

Not applicable

Set 2 follows these
parameters

See next page

(1)

Up to 2 decimal places on Series 3L

Set 1 is followed

Set this for the maximum

with R N G . H I

display range required

Then

Set this for the minimum

R N G . L o

display range required

Home display

N

PV only

A

First Alarm SP only

1

PV + Alarm SP

2

PV + Alarm SP

(Read only)