9.4 Diagnostic Alarms

Diagnostic alarms indicate a possible fault within the indicator or connected devices.

Display shows

What it means

What to do about it

E.Conf

A change made to a parameter takes a finite time
to be entered. If the power to the indicator is
turned off before the change has been entered
then this alarm will occur.

Do not turn the power off to the indicator while
ConF is flashing

Enter configuration mode then return to the
required operating mode. It may be necessary to
re-enter the parameter change since it will not have
been entered in the previous configuration.

E.CaL

Calibration error

Re-instate Factory calibration

E2.Er

EEPROM error

Return to factory for repair

EE.Er

Non-vol memory error

Note the error and contact your supplier

E.Lin

Invalid input type. This refers to custom
linearisation which may not have been applied
correctly or may have been corrupted.

Go to the INPUT list in configuration level and set a
valid thermocouple or input type