4 to check rtd input calibration, 2 offsets, 1 five point offset – Super Systems 3L Series User Manual

Page 47: To check rtd input calibration, Offsets, Five point offset, N 12.2. f

Advertising
background image

Series 3L Indicators

47

12.1.4 To Check RTD Input Calibration

Connect a decade box with total resistance lower than
1K and resolution to two decimal places in place of the
RTD as indicated on the connection diagram below
before the instrument is powered up. If at any
instant the instrument was powered up without this
connection then at least 10 minutes must elapse from
the time of restoring this connection before RTD
calibration check can take place.





The RTD range of the instrument is -200 to 850

O

C. It

is, however, unlikely that it will be necessary to check
the instrument over this full range.

Set the resistance of the decade box to the minimum
range. For example 0

O

C = 100.00

Ω. Check the

calibration is within +0.25% of 0

O

+ 1LSD.

Set the resistance of the decade box to the maximum
range. For example 200

O

C = 175.86

Ω. Check the

calibration is within +0.25% of 200

O

+ 1LSD.

12.2 Offsets

The process value can be offset to take into account
known errors within the process. The offset can be
applied to any Input Type (mV, V, mA, thermocouple or
RTD).

A single offset can be applied - the procedure is carried
out in the

INPUT list and has been described in section

6.1.3.

It is also possible to adjust the low and high points as a
five point offset. This can only be done in Level 3 in the

Cal’ list and is described below.

12.2.1 Five Point Offset

A five point offset may be used to compensate for
transducer or measurement non-linearities. The
diagram shows an example of the type of discontinuity
which might occur in a system.

In this case adjust each point in turn for the VALUE
WHICH THE INDICATOR SHOULD READ. For
example if the value at point 1 should be 1.2345 then
set

Pnt.1 to this value. The following example shows

how to do this.

Do This

Display View

Notes

1.

Select Level 3 as

described in Chapter 2.

Then press

to

select ‘CAL’

Scrolling
message

c a l i b r a t i o n
l i s t

2.

Press

to scroll

to ‘u.cal’

Scrolling
message

u s e r
c a l i b r a t i o n

3.

Press

or

to ‘

pnt.1’

To revert to
the original
values,
select

rSet

4.

Press

to scroll

to ‘C.Adj’

5.

Press

or

to the correct value

Note:- this is
not an offset
value

6.

Repeat the above

for points 1 to 5

In some cases it will not be necessary to adjust all 5
points. For example, a low and high adjustment may be
all that is necessary as shown in the following
diagrams.

In this case set

Pnt.1 to the required low point value.

For the high point value you may select any point

Pnt.2 to Pnt.5. The instrument applies a straight line
between the two points.

Note:-

The calibration points must be chosen consecutively –
the five point calibration will not work if a higher point is
inserted between other points.

Electrical
Input

Display
Reading

High offset

(e.g. 6)

Factory
calibration

Low offset

(e.g. 1)

Electrical
Input

Display
Reading

High offset

Factory
calibration

Low offset

Output Hi

Output Lo

Input Hi

Input Lo

Cal Point 5

Cal Point 4

Cal Point 3

Cal Point 2

Cal Point 1

Matched impedance copper

leads

Decade

Box

Indicator

VI

V-

V+

Advertising