3 input calibration, 1 to calibrate mv range, 2 to calibrate thermocouple ranges – Super Systems 3L Series User Manual

Page 48: Input calibration, To calibrate mv range, To calibrate thermocouple ranges

Advertising
background image

Operations Manual

Series 3L

48

12.3 Input Calibration

If the calibration is not within the specified accuracy
follow the procedures in this section:

In Series 3L series instruments, inputs which can be
calibrated are:

mV Input. This is a linear 80mV range calibrated
at two fixed points. This should always be done
before calibrating either thermocouple or resistance
thermometer inputs. mA range calibration is
included in the mV range.

Thermocouple calibration involves calibrating the
temperature offset of the CJC sensor only. Other
aspects of thermocouple calibration are also
included in mV calibration.

Resistance Thermometer. This is also carried out
at two fixed points - 150

Ω and 400Ω.

12.3.1 To Calibrate mV Range

Calibration of the mV range is carried out using a 50
milli-volt source, connected as shown in the diagram
below. mA calibration is included in this procedure.

For best results 0mV should be calibrated by
disconnecting the copper wires from the mV source and
short circuiting the input to the indicator

Select Conf Level as described in Chapter 2, s

et the

indicator input to mV range, then:-

Do This

Display View

Additional Notes

1.

From any

display press

as many times as
necessary until the
‘CAL’ page header
is displayed.

Scrolling display

C A L I B R A T I O N

L I S T ’

2.

Press

to

select ‘P H A S E ’

none

p h a s e

Scrolling display

C A L I B R A T I O N

p h a s e ’

3.

Set mV source for 0mV

4.

Press

or

to choose ‘

0’

5.

Press

to

select ‘G O ’

6.

Press

or

to choose

YES’

Scrolling display

C A L I B R A T I O N

s t a r t ’

The indicator
automatically
calibrates to the
injected input mV.

As it calibrates the display will show

busy

then

pass,

assuming a

successful calibration.

If it is not successful then ‘

FAIL

’ will

be displayed. This may be due to
incorrect input mV

7.

Set mV source for 50mV

8.

Press

to

select ‘P H A S E ’

9.

Press

or

to choose ‘

50’

10. Repeat 5 & 6
above

The indicator
calibrates to the
high point in the
same way as the
low point above

12.3.2 To Calibrate Thermocouple Ranges

Thermocouples are calibrated, firstly, by following the
previous procedure for the mV ranges, then calibrating
the CJC.

This can be carried out using an external CJC reference
source such as an ice bath or using a thermocouple mV
source. Replace the copper cable shown in the
diagram below with the appropriate compensating cable
for the thermocouple in use.

Set the mV source to internal compensation for the
thermocouple in use and set the output for 0mV.
Then:-

Do This

Display View

Additional

Notes

1.

From the mV

calibration, press

or

to select

CJC’

2.

Press

to

select ‘GO’

3.

Press

or

to choose ‘

YES’

The indicator
automatically
calibrates to
the CJC input
at 0mV.

As it does this
the display will
show

busy

then

pass

,

assuming a
successful
calibration.
If it is not
successful
then ‘

FAIL

’ will

be displayed.
This may be
due to an
incorrect input
mV

Copper cable

50 mV

Source

+

-

Indicator VI

V-

V+

Thermocouple
Compensating

cable

Thermocouple

simulator set to

T/C type and 0

o

C

+

-

Controller

VI

V-

V+

Advertising