3 to calibrate rtd ranges, To calibrate rtd ranges – Super Systems 3L Series User Manual

Page 49

Advertising
background image

Series 3L Indicators

49

12.3.3 To Calibrate RTD Ranges

The two points at which the RTD range is calibrated are
150.00

Ω and 400.00Ω.

Before starting RTD calibration:

A decade box with total resistance lower than 1K
must be connected in place of the RTD as
indicated on the connection diagram below before
the instrument is powered up
. If at any instant
the instrument was powered up without this
connection then at least 10 minutes must elapse
from the time of restoring this connection before
RTD calibration can take place.

The instrument should be powered up for at least
10 minutes.

Before using or verifying RTD calibration:

The mV range must be calibrated first.

Do This

Display View

Notes

1.

From any display

press

as many

times as necessary until
the ‘C A L ’ page header
is displayed.

Scrolling
display

C A L I B R A

T I O N
L I S T ’

2.

Press

to select

‘P H A S E ’

Scrolling
display

C A L I B R A

T I O N
p h a s e ’

3.

Set the decade box for 150.00

4.

Press

or

to choose ‘150r’’

5.

Press

to select

‘GO’

6.

Press

or

to choose ‘

YES’

Scrolling
display

C A L I B R A

T I O N
s t a r t ’

The indicator automatically calibrates to the injected
150.00

Ω input.

As it does this the display will show

busy

then

pass

,

assuming a successful calibration.

If it is not successful then ‘

FAIL

’ will be displayed. This

may be due to an incorrect input resistance

7.

Set the decade box for 400.00

8.

Press

or

to choose ‘400r’’

9.

Repeat 5 and 6

above to calibrate the
high point

The indicator will again automatically calibrate to the
injected 400.00

Ω input.

If it is not successful then ‘

FAIL

’ will be displayed

Matched

impedance

copper leads

Decade Box

Controller

VI

V-

V+

Advertising