Super Systems 3L Series User Manual

Page 44

Advertising
background image

Operations Manual

Series 3L

44

– others off
1 – All outputs inactive

uCAL

User Calibration Enable

533

A1.TYP





Alarm 1 Type
0 – Off
1 –Absolute High
2 – Absolute Low
3 – Deviation High
4 – Deviation Low
5 – Deviation Band

536





A2.TYP

Alarm 2 Type
(as Alarm 1 Type)

537

A3.TYP

Alarm 3 Type
(as Alarm 1 Type)

538

A4.TYP

Alarm 4 Type
(as Alarm 1 Type)

539

A1.LAT


Alarm 1 Latching Mode
0 – No latching
1 – Latch - Automatic Reset
2 – Latch – Manual Reset

540


A2.LAT

Alarm 2 Latching Mode
(as Alarm 1 Latching Mode)

541

A3.LAT

Alarm 3 Latching Mode
(as Alarm 1 Latching Mode)

542

A4.LAT

Alarm 4 Latching Mode
(as Alarm 1 Latching Mode)

543

A1.BLK

Alarm Blocking Mode Enable (0 =
OFF, 1 = BLOCK)

544

A2.BLK

Alarm Blocking Mode Enable (0 =
OFF, 1 = BLOCK)

545

A3.BLK

Alarm Blocking Mode Enable (0 =
OFF, 1 = BLOCK)

546

A4.BLK

Alarm Blocking Mode Enable (0 =
OFF, 1 = BLOCK)

547

Di.OP






Digital Outputs Status. This is a
bitmap:
B0 – Output 1A
B1 – (not used)
B2 – Output 3A
B3 – Output 4/AA
It is possible to write to this status
word to use the digital outputs in a
telemetry output mode. Only
outputs whose function is set to
‘none’ are affected, and the setting
of any bits in the Digital Output
Status word will not affect outputs
used for heat (for example) or other
functions. Thus it is not necessary
to mask in the settings of these bits
when writing to this variable.

551






Alarm 1 delay

552

Alarm 2 delay

553

Alarm 3 delay

554

Alarm 4 delay

555

Alarm 1 offset

556

Alarm 2 offset

557

Alarm 3 offset

558

Alarm 4 offset

559

OFS.HI

Adjust High Offset

560

OFS.LO

Adjust Low Offset

561

PNT.HI

Adjust High Point

562

PNT.LO

Adjust Low Point

563

SB.TYP

Sensor Break Type
0 – No Sensor Break
1 – Non-Latching Sensor Break

578

2 – Latching Sensor Break

SB.DIR

Sensor break direction
Up –
Down –

579

Id

Customer ID – May be set to any
value between 0-9999 for
identification of instruments in
applications. Not used by the
instrument itself.

629

PHASE








Calibration Phase
0 – None
1 – 0 mv
2 – 50 mv
3 – 150 Ohm
4 – 400 Ohm
5 – CJC
6 – CT 0 mA
7 – CT 70 mA
8 – Factory Defaults

768








GO




Calibration Start
0 – No
1 – Yes (start cal)
2 – Cal Busy
3 – Cal Pass
4 – Cal Fail
Note values 2-4 cannot be written
but are status returns only

769




K.LOC

Allows instrument to be locked via a
key/digital input
0 - unlocked,
1 – all keys locked
2 – Edit keys (raise and lower)
disabled
3 – Mode key disabled
4 – Manual mode disabled
5 – Enter standby mode when
Mode combination pressed
6 – Timer keys disabled

1104

IN.TYP











Input Sensor Type
0 – J Type Thermocouple
1 – K Type Thermocouple
2 – L Type Thermocouple
3 – R Type Thermocouple
4 – B Type Thermocouple
5 – N Type Thermocouple
6 – T Type Thermocouple
7 – S Type Thermocouple
8 – RTD
9 – millivolt
10 – Comms Input (see Modbus
address 203)
11 – Custom Input (Downloadable)

12290











CJ.TYP

CJC Type
0 – Auto
1 – 0 Degrees C
2- 50 Degrees C

12291

mV.HI

Linear Input High

12306

mV.LO

Linear Input Low

12307

L.TYPE

Logic Input A channel hardware
type
0 – None
1 – Logic Inputs

12352

L.D.IN





Logic input A function
40 – None
41 – Acknowledge all alarms
42 – Select SP1/2
43 – Lock All Keys
44 – Timer Reset
45 – Timer Run

12353





Advertising