Source[<n>]][:mod]:am:source – RIGOL DG1000Z Series User Manual

Page 132

Advertising
background image

RIGOL

Chapter 2 Command System

2-120

DG1000Z Programming Guide

[:SOURce[<n>]][:MOD]:AM:SOURce

Syntax

[:SOURce[<n>]][:MOD]:AM:SOURce {INTernal|EXTernal}

[:SOURce[<n>]][:MOD]:AM:SOURce?

Description

Set the AM modulation source of the specified channel to internal (INTernal) or

external (EXTernal) modulation source.
Query the AM modulation source of the specified channel.

Parameter

Name

Type

Range

Default

[<n>]

Discrete

1|2

1

{INTernal|EXTernal}

Discrete

INTernal|EXTernal

INTernal

Explanation

 DG1000Z can receive the modulation waveform from the internal or external

modulation source.

 When internal modulation source is selected, the modulation waveform can be

SINusoid, SQUare, TRIangle, RAMP, NRAMp, NOISe or USER and the default is

SINusoid. NOISe can be used as modulation waveform but cannot be used as

carrier waveform.

 When external modulation source is selected, the signal generator receives the

external modulating signal from the [Mod/Trig/FSK/Sync] connector of the

corresponding channel at the rear panel. At this point, the amplitude of the

modulated waveform is controlled by the ±5 V signal level of the connector. For

example, if the modulation depth is set to 100%, the output amplitude will be

the maximum when the modulating signal is +5V and the minimum when the

modulating signal is -5V.

 When [:SOURce[<n>]] or [<n>] is omitted, the system sets the related

parameters of CH1 by default.

Return Format

The query returns INT or EXT.

Example

:SOUR1:AM:SOUR EXT /*Set the AM modulation source of CH1 to external

modulation source*/

:SOUR1:AM:SOUR? /*Query the AM modulation source of CH1 and the query

returns EXT*/