Chapter 2 Command System

RIGOL

DG1000Z Programming Guide

2-139

[:SOURce[<n>]][:MOD]:PM:SOURce

Syntax

[:SOURce[<n>]][:MOD]:PM:SOURce {INTernal|EXTernal}
[:SOURce[<n>]][:MOD]:PM:SOURce?

Description

Set the PM modulation source of the specified channel to internal (INTernal) or

external (EXTernal) modulation source.
Query the PM modulation source of the specified channel.

Parameter

Name

Type

Range

Default

[<n>]

Discrete

1|2

1

{INTernal|EXTernal}

Discrete

INTernal|EXTernal

INTernal

Explanation

 DG1000Z can receive the modulation waveform from the internal or external

modulation source.

 When internal modulation source is selected, the modulation waveform can be

SINusoid, SQUare, TRIangle, RAMP, NRAMp, NOISe or USER and the default is

SINusoid. NOISe can be used as modulation waveform but cannot be used as

carrier waveform.

 When external modulation source is selected, the signal generator receives the

external modulating signal from the corresponding [Mod/Trig/FSK/Sync]

connector at the rear panel. At this point, the phase deviation of the

modulated waveform is controlled by the ±5 V signal level of the connector.

For example, if the phase deviation is set to 180°, +5V signal level

corresponds to a 180º phase variation. The lower the external signal level, the

less deviation would be generated.

 When [:SOURce[<n>]] or [<n>] is omitted, the system sets the related

parameters of CH1 by default.

Return Format

The query returns INT or EXT.

Example

:SOUR1:PM:SOUR EXT /*Set the PM modulation source of CH1 to external

modulation source*/

:SOUR1:PM:SOUR? /*Query the FM modulation source of CH1 and the

query returns EXT*/

Advertising