Coupling:ampl:deviation – RIGOL DG1000Z Series User Manual

Page 22

Advertising
background image

RIGOL

Chapter 2 Command System

2-10

DG1000Z Programming Guide

:COUPling:AMPL:DEViation

Syntax

:COUPling:AMPL:DEViation <deviation>

:COUPling:AMPL:DEViation?

Description

Set the amplitude deviation in the amplitude coupling.
Query the amplitude deviation in the amplitude coupling.

Parameter

Name

Type

Range

Default

<deviation>

Real

-19.998Vpp to 19.998Vpp

0Vpp

Explanation

 Select the desired amplitude coupling mode (

:COUPling:AMPL:MODE

) and set

the corresponding amplitude deviation or amplitude ratio

(

:COUPling:AMPL:RATio

) before enabling the amplitude coupling function

(

:COUPling:AMPL[:STATe]

). You cannot set the amplitude coupling mode and

amplitude deviation/ratio after the amplitude coupling function is enabled.

 When the amplitude coupling function is disabled, if the current amplitude

coupling mode is amplitude deviation, send this command can set the amplitude

deviation; if the current amplitude coupling mode is amplitude ratio, send this

command can set the amplitude coupling mode to amplitude deviation and set

the amplitude deviation.

Return Format

The query returns the amplitude deviation in scientific notation. The return value

contains 7 effective digits, for example, 1.000000E+00 (the amplitude deviation is

1Vpp).

Example

:COUP:AMPL:DEV 1 /*Set the amplitude deviation in the amplitude coupling to

1Vpp*/

:COUP:AMPL:DEV? /*Query the amplitude deviation in the amplitude coupling

and the query returns 1.000000E+00*/

Related

Commands

:COUPling:AMPL:MODE
:COUPling:AMPL:RATio

:COUPling:AMPL[:STATe]