[:SOURce[<n>]]:APPLy:USER

Syntax

[:SOURce[<n>]]:APPLy:USER

[<freq>|DEFault|MINimum|MAXimum[,<amp>|DEFault|MINimum|MAXimum

[,<offset>|DEFault|MINimum|MAXimum[,<phase>|DEFault|MINimum|MAXimum]]]]

Description

Set the waveform of the specified channel to arbitrary waveform (in frequency output

mode) with the specified frequency, amplitude, offset and phase.

Parameter

Name

Type

Range

Default

[<n>]

Discrete

1|2

1

<freq>

Real

1uHz to 20MHz

1kHz

<amp>

Real

Refer to the “Explanation”

5Vpp

<offset>

Real

Refer to the “Explanation”

0V

DC

<phase>

Real

0° to 360°

Explanation

 When [:SOURce[<n>]] or [<n>] is omitted, the system sets the related

parameters of CH1 by default.

 This command only sets the arbitrary waveform parameters of the channel and

does not set the arbitrary waveform type. The default arbitrary waveform is Sinc.

You can send the

[:SOURce[<n>]]:FUNCtion[:SHAPe]

command to set the

waveform of the specified channel to the desired arbitrary waveform.

 The range of <amp> is limited by the “Impedance” (

:OUTPut[<n>]:IMPedance

or

:OUTPut[<n>]:LOAD

) and “Frequency/Period” settings. The range of <offset>

is limited by the “Impedance” and “Amplitude/High Level” settings.

Example

:SOUR1:APPL:USER 100,1,2,3 /*Set the waveform of CH1 to arbitrary waveform

(in frequency output mode) with 100Hz

frequency, 1Vpp amplitude, 2V

DC

offset and 3°

start phase*/