System:communicate:gpib[:self]:address, System:communicate:lan:apply – RIGOL DG1000Z Series User Manual

Page 201

Advertising
background image

Chapter 2 Command System

RIGOL

DG1000Z Programming Guide

2-189

:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess

Syntax

:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess <integer>

:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess?

Description

Set the GPIB address of the instrument.
Query the GPIB address of the instrument.

Parameter

Name

Type

Range

Default

<integer>

Integer

0 to 30

2

Explanation

To use the GPIB interface, make sure that your PC has installed the GPIB card. Then,

connect the USB terminal of the USB-GPIB interface converter to the USB Host

interface at the front panel of the signal generator and connect the GPIB terminal of

the USB-GPIB interface converter to the GPIB card terminal of the PC.

Return Format

The query returns the GPIB address in scientific notation with 7 effective digits, for

example, 7.000000E+00 (the GPIB address is 7).

Example

:SYST:COMM:GPIB:ADDR 7 /*Set the GPIB address of the instrument to 7*/
:SYST:COMM:GPIB:ADDR? /*Query the GPIB address of the instrument and

the query returns 7.000000E+00*/

:SYSTem:COMMunicate:LAN:APPLy

Syntax

:SYSTem:COMMunicate:LAN:APPLy

Description

Apply the current network parameters.

Explanation

After setting the LAN parameters, the new settings will only take effect when this

command is executed.