:SYSTem:BEEPer[:IMMediate]

Syntax

:SYSTem:BEEPer[:IMMediate]

Description

The beeper generates a beep immediately.

Explanation

This command will not be affected by the current on/off status of the beeper. The

beeper will generate a beep immediately when this command is sent even if the

beeper is currently turned off.

:SYSTem:BEEPer:STATe

Syntax

:SYSTem:BEEPer:STATe {ON|1|OFF|0}

:SYSTem:BEEPer:STATe?

Description

Turn on or off the beeper.
Query the beeper status.

Parameter

Name

Type

Range

Default

{ON|1|OFF|0}

Bool

ON|1|OFF|0

ON

Explanation

When the beeper is turned on, the beeper beeps when error occurs during

front-panel operation or remote operation.

Return Format

The query returns ON or OFF.

Example

:SYST:BEEP:STAT 1 /*Turn on the beeper */
:SYST:BEEP:STAT? /*Query the beeper status and the query returns ON*/

:SYSTem:CHANnel:CURrent

Syntax

:SYSTem:CHANnel:CURrent {CH1|CH2}

:SYSTem:CHANnel:CURrent?

Description

Select the current channel.
Query the current channel.

Parameter

Name

Type

Range

Default

{CH1|CH2}

Discrete

CH1|CH2

CH1

Return Format

The query returns CH1 or CH2.

Example

:SYST:CHAN:CUR CH2 /*Select CH2 as the current channel*/
:SYST:CHAN:CUR? /*Query the current channel and the query returns

CH2*/

:SYSTem:CHANnel:NUMber?

Syntax

:SYSTem:CHANnel:NUMber?

Description

Query the number of channels of the instrument.

Return Format

The query returns an integer.

Example

:SYST:CHAN:NUM? /*Query the number of channels of the instrument and the

query returns 2*/