Counter:hf, Counter:level – RIGOL DG1000Z Series User Manual

Page 17

Advertising
background image

Chapter 2 Command System

RIGOL

DG1000Z Programming Guide

2-5

:COUNter:HF

Syntax

:COUNter:HF {ON|1|OFF|0}

:COUNter:HF?

Description

Enable or disable the high-frequency rejection function of the frequency counter.
Query the on/off status of the high-frequency rejection function of the frequency

counter.

Parameter

Name

Type

Range

Default

{ON|1|OFF|0}

Bool

ON|1|OFF|0

OFF

Explanation

Enable the high-frequency rejection when measuring low-frequency signal with

lower than 250kHz frequency to filter out the high-frequency noise and improve

the measurement accuracy. Disable the high-frequency rejection when measuring

high-frequency signal with greater than 250kHz frequency; at this point, the

maximum input frequency is 200MHz.

Return Format

The query returns ON or OFF.

Example

:COUN:HF ON /*Enable the high-frequency rejection function of the frequency

counter*/

:COUN:HF? /*Query the on/off status of the high-frequency rejection

function of the frequency counter and the query returns ON*/

:COUNter:LEVEl

Syntax

:COUNter:LEVEl {<value>|MINimum|MAXimum}

:COUNter:LEVEl? [MINimum|MAXimum]

Description

Set the trigger level of the frequency counter.
Query the trigger level of the frequency counter.

Parameter

Name

Type

Range

Default

<value>

Real

-2.5V to 2.5V

0V

Explanation

 The frequency counter starts measuring when the input signal reaches the

specified trigger level.

 The minimum resolution is 6mV.

Return Format

The query returns the trigger level in scientific notation. The return value contains 7

effective digits, for example, 1.500000E+00 (the trigger level is 1.5V).

Example

:COUN:LEVE 1.5 /*Set the trigger level of the frequency counter to 1.5V*/
:COUN:LEVE? /*Query the trigger level of the frequency counter and the

query returns 1.500000E+00*/