Output[<n>]:sync:polarity – RIGOL DG1000Z Series User Manual

Page 72

Advertising
background image

RIGOL

Chapter 2 Command System

2-60

DG1000Z Programming Guide

:OUTPut[<n>]:SYNC:POLarity

Syntax

:OUTPut[<n>]:SYNC:POLarity {POSitive|NEGative}

:OUTPut[<n>]:SYNC:POLarity?

Description

Set the output polarity of the sync signal on the [Mod/Trig/FSK/Sync] connector

of the specified channel at the rear panel to normal (POSitive) or inverted

(NEGative).
Query the output polarity of the sync signal on the [Mod/Trig/FSK/Sync]

connector of the specified channel at the rear panel.

Parameter

Name

Type

Range

Default

[<n>]

Discrete

1|2

1

{POSitive|NEGative}

Discrete

POSitive|NEGative

POSitive

Explanation

 When [<n>] is omitted, the commands set and query the related parameters of

CH1 by default.

 The output polarity of the sync signal refers to that the sync signal on the

[Mod/Trig/FSK/Sync] connector of the channel at the rear panel is normal

(POSitive) output or inverted (NEGative) output. In the normal mode, the

instrument outputs the sync signal normally and in the inverted mode, the sync

signal is inverted and then outputted.

 After the waveform is inverted (

:OUTPut[<n>]:POLarity

), the sync signal related

to the waveform will not be inverted.

Return Format

The query returns POS or NEG.

Example

:OUTP1:SYNC:POL POS /*Set the output polarity of the sync signal on the

[Mod/Trig/FSK/Sync] connector of CH1 at the rear

panel to normal */

:OUTP1:SYNC:POL? /*Query the output polarity of the sync signal on the

[Mod/Trig/FSK/Sync] connector of CH1 at the rear

panel and the query returns POS*/

Related

Command

:OUTPut[<n>]:POLarity

Advertising