System:roscillator:source, System:security:immediate – RIGOL DG1000Z Series User Manual

Page 213

Advertising
background image

Chapter 2 Command System

RIGOL

DG1000Z Programming Guide

2-201

:SYSTem:ROSCillator:SOURce

Syntax

:SYSTem:ROSCillator:SOURce {INTernal|EXTernal}

:SYSTem:ROSCillator:SOURce?

Description

Set the system clock source to internal source (INTernal) or external source

(EXTernal).
Query the system clock source type.

Parameter

Name

Type

Range

Default

{INTernal|EXTernal}

Discrete

INTernal|EXTernal

INTernal

Explanation

 DG1000Z provides an internal 10MHz clock source and accepts external clock

source from the [10MHz In/Out] connector at the rear panel. It can also

output clock source from the [10MHz In/Out] connector for other devices.

 If external clock source is selected, the instrument will detect whether a valid

external clock signal is inputted from the [10MHz Out/In] connector at the

rear panel. If no valid clock source is detected, the corresponding prompt

message would be displayed and the clock source would be switched to

internal.

 You can synchronize two or more instruments by setting the clock source.

Synchronization between two instruments:

Connect the output of the [10MHz In/Out] connector of instrument A (set

the clock source to “Internal”) to the [10MHz In/Out] connector of

instrument B (set the clock source to “External”) and set the output

frequencies of instruments A and B as a same value to realize synchronization

between the two instruments.
Synchronization among multiple instruments:

Divide the 10MHz clock source of a instrument (set the clock source to

“Internal”) into multiple instruments, and then connect them to the [10MHz

In/Out] connectors of other generators (set the clock source to “External”)

respectively, and finally set the output frequencies of all the instruments as a

same value to realize synchronization among the multiple instruments.

Return Format

The query returns INT or EXT.

Example

:SYST:ROSC:SOUR INT /*Set the system clock source to internal*/
:SYST:ROSC:SOUR? /*Query the system clock source and the query returns

INT*/

:SYSTem:SECurity:IMMediate

Syntax

:SYSTem:SECurity:IMMediate

Description

Sanitizes all the user-accessible instrument memory information.

Explanation

This command sanitizes all the user-defined state information, arbitrary waveforms

and I/O settings (such as the IP address) as well as restores the instrument settings

to the factory values.

Advertising