Chapter 2 Command System

RIGOL

DG1000Z Programming Guide

2-7

:COUNter[:STATe]

Syntax

:COUNter[:STATe] {ON|1|OFF|0|RUN|STOP|SINGLE}

:COUNter[:STATe]?

Description

Set the status of the frequency counter.
Query the status of the frequency counter.

Parameter

Name

Type

Range

Default

{ON|1|OFF|0|RUN|STOP|

SINGLE}

Discrete

ON|1|OFF|0|RUN|STOP|

SINGLE

OFF

Explanation

 “ON” and “1” denote enabling the frequency counter function; “OFF” and “0”

denote disabling the frequency counter function; “RUN”, “STOP” and “SINGLE”

denote setting the running status of the frequency counter to “run”, “stop” and

“single” respectively.

 The command for setting the running status (the parameter is RUN, STOP or

SINGLE) is only valid when the frequency counter function is enabled.

 When the frequency counter function is enabled, the sync output of CH2 will be

disabled.

 In the “RUN” status, the frequency counter measures the input signal

continuously according to the current configuration. In the “SINGLE” status, the

frequency counter executes a measurement and then stops. In the “STOP”

status, the frequency counter stops measuring.

 When the frequency counter is enabled, the default running status is “run” and

the instrument measures the input signal continuously according to the current

configuration. At this point, if you send the :COUNter:STATe SINGLE command,

the frequency counter enters the “single” status, finishes the current

measurement and then stops; if you send the :COUNter:STATe STOP command,

the frequency counter enters the “STOP” state immediately.

 When the frequency counter is in the “STOP” status, the frequency counter

performs a measurement and then enters the “STOP” status each time you send

the :COUNter:STATe SINGLE command.

Return Format

When the frequency counter function is enabled, the query returns the current

running status (RUN, STOP or SINGLE); when the frequency counter function is

disabled, the query returns OFF.

Example

:COUN OFF /*Disable the frequency counter function*/
:COUN? /*Query the status of the frequency counter and the query returns

OFF*/

:COUN 1 /*Enable the frequency counter function*/
:COUN? /*Query the status of the frequency counter and the query returns

RUN (the default running status)*/

:COUN STOP /*Set the running status of the frequency counter to “STOP”*/
:COUN? /*Query the status of the frequency counter and the query returns

STOP*/