Appendix – Gilson 215 User Manual

Page 102

Advertising
background image

C-4

Appendix

C

GSIOC Commands

9

B

Clears the contact event FIFO and resets the contact input

event timer.

A

I

Reads home phases for X and Y motors. Returns “xxx/yyy”

where:

xxx - X motor phase.

yyy - Y motor phase.

Bf,d

B

Causes 215 to beep:

f - Frequency in Hz; default is 2400.

d - Duration of beep in tenths of seconds; default is 1.

cv=ss..ss

B

Sets user-definable character:

v

- Address (0 to 7) of user-defined character.

ss..ss - Characters, in hexadecimal ASCII code, for

rows 1 through 7. See example below.

For example, the following command sets the pattern

shown below to user-defined character address 0:

c0=1F111B151B111

[Row]

[Pattern]

[Hex Code]

1

* * * * *

1F

2

* *

11

3

* * * *

1B

4

* * *

15

5

* * * *

1B

6

* *

11

7

* * * * *

1F

e

I

Reads the current error number. Returns “nnn” which

identifies the error number; see page

5-2

for listing of

errors. If no error has occurred, returns 0.

e

B

Clears error number.

Note: Whenever possible, it is recommended that the

immediate “$” command be used to clear errors.

Advertising