Replacing a Level Sensing Cable

Removing the level sensing cable

1 Detach the transfer tubing

from the top of the isolation

probe holder by using the

supplied headless nut extender

(part number 49041032) to

loosen and remove the 1/4"-28

nut and ferrule.

2 Unplug the right-angle

connector for the liquid level

sensing cable (LLD cable) from

the socket on the back of the

Z-arm.

3 Detach the LLD cable from the

LLD cable guide.

4 Loosen the small hexagonal

nut on the probe holder for

probe one. Detach the split-

tongue connector from the nut.

5 Remove the transfer tubing

from the LLD cable casing

.

Replacing Parts