Gilson 215 User Manual

Page 60

Advertising
background image

4-4

Maintenance

4

Cleaning

Cleaning methods

Depending on the samples or reagents that come

into contact with the fluid path, you may need to

vary your cleaning methods accordingly. Use the

following cleaning protocols as references and make

any changes to them as required for the samples and

reagents being pumped for your application.

Proteins and peptides - Follow this procedure if

the fluid path is in contact with proteins and

peptides:

1 Prime the fluid path using a weak detergent

solution.

2 Pause the priming sequence.

3 After 30 minutes, resume priming the fluid path

using distilled or deionized water to pump the

remaining detergent from the tubing into a waste

container. Prime the fluid path a minimum of 10

cycles with distilled or deionized water.

4 When you’re satisfied that the entire fluid path

has been flushed with water, end the priming

sequence.

Advertising