4-5

Maintenance

4

Cleaning

Acidic and basic compounds - Follow this

procedure if the fluid path is in contact with acidic

and basic compounds:

1 Prime the fluid path using a 0.1N NaOH

solution.

2 Pause the priming sequence.

3 After 10 minutes, resume priming the fluid path

using distilled or deionized water. Prime until

the fluid path has been flushed with water.

4 Pause the priming sequence.

5 Prime the fluid path using a 0.1N NaOH

solution. Continue to prime until the fluid path

has been flushed with 0.1N NaOH.

6 Pause the priming sequence.

7 After 10 minutes, resume priming the fluid path

using distilled or deionized water. Prime until

the fluid path has been flushed with water.

8 When you’re satisfied that the entire fluid path

has been flushed with water, end the priming

sequence.