3-10

Operation

3

Configuring the Liquid Handler

Adjust XY

The Adjust XY tab allows you to test whether the

instrument is properly adjusted and to make minor

adjustments to the X- and Y-axis offsets if needed.

You may need to use the options under this tab if

the probe is not accessing the injection port of the

injection module installed or the vessels in the

installed racks.

The X offset and Y offset text boxes display the

current offsets stored in the instrument’s memory.

To determine if the probe(s) need(s) to be adjusted

in the X or Y direction select the model of the

injection module that is on your Quad-Z 215 or

select other for a user-defined test point. The

default XY coordinates are shown next to the

model number.

Select the model of the injection module that is on

your 215 or select other for a user-defined test

point. The default XY coordinates are shown next

to the model number.