Maintenance

4

Replacing Parts

Replacing Tubing

It is important to keep all tubing clean and free of

crimps. Tubing that has become dirty, blocked or

crimped can result in poor accuracy and precision,

or loss of air gap.

Replace both the transfer tubing and inlet tubing as

needed. See

Appendix A

for part numbers for

replacement tubing. For tubing installation

procedures, see

Section 2

.

Replacing a Probe

Refer to the appropriate instructions below

depending on whether you’re replacing a probe

with one of the same type or one of a different

type.

Installing same type of probe

To install a replacement probe of

the same type that’s currently

installed:

1 Remove the transfer tubing’s

1/4"-28 fitting connected to the

top of the isolation probe

holder.

2 Grasp the current probe and

push it up through the top of

the isolation probe holder.

3 Insert the probe into the top of

the isolation probe holder and

pull it through the holder and

the probe guide insert until the

tip of the probe is in the probe

guide.

4 Replace and tighten the 1/4"-28

fitting.

Replacing Parts