Rinse site, Rinse site -8 – Gilson 215 User Manual

Page 48

Advertising
background image

3-8

Operation

3

Configuring the Liquid Handler

Rinse Site

The instrument stores the location of a rinse site in

memory. This allows the instrument to move to this

location before homing the syringe and prevents

the spilling of waste liquid or rinse diluent.

You must modify the rinse site coordinates if you

have a Quad-Z 215 Liquid Handler. Refer to the

following table to select the correct X, Y, and Z

coordinates depending on the kind of probe rinse

that will be performed. Coordinates for the rinse

site are provided for the X, Y, and Z axis in

millimeters. The coordinates identify the rinse site

for the left-most probe installed on the Quad-Z 215

Liquid Handler.

X

Y

Z

Shallow-pocket rinse

82.9

16.5

100.0

Deep-pocket rinse

82.9

3.8

47.5

Flow-through rinse

82.9

3.8

47.5