Helpful Hints

Helpful Hints

In order to keep your liquid handler at peak

performance, Gilson recommends that you do the

following:

• Change or clean the tubing regularly to

maintain maximum performance.

• Flush the probe housings and rinse stations

daily with distilled or deionized water. On a

weekly basis, flush with a 10% solution of

bleach or weak detergent.

• Check periodically to ensure that all fittings are

tight.

• Wipe up all spills immediately.

• Cold fluids may cause leakage; warm fluids to

room temperature before running them through

the system.

• Lubricate the rods on the Z-arm at least once

every six months (see page