Armlock removal, Armlock removal -3 – Gilson 215 User Manual

Page 19

Advertising
background image

2-3

Installation

2

Armlock Removal

Armlock Removal

The armlock on the liquid handler secures the Y-arm

during shipment. You must remove the armlock

prior to installing the Z-arm and operating the

instrument. If the armlock is not removed, the

liquid handler cannot move in the X-direction. This

results in an error state during operation.

If you need to move the liquid handler, always

reinstall the armlock. This safeguards against

mechanical damage.

To remove the armlock:

1 Remove the cardboard label in front of the

armlock.

2 Using the 9/64" ball driver, loosen the

tensioning screw that immobilizes the Y-arm.

3 Using the 9/64" ball driver, remove the two

remaining screws that hold the armlock in

place.

4 Remove the armlock and store it and the ball

driver for future use.