2-15

Installation

2

4 Loosen the mounting screw on the Z-arm

mounting bracket and slightly raise the Z-arm.

5 Manually move the liquid handler’s arm so the

Z-arm is centered over the Z-height adjustment

tool.

6 While holding Z-height adjustment tool flat

against the locator plate, use the other hand to

lower the Z-arm until it lightly rests on the

adjustment tool.

7 Tighten the mounting screw on the Z-arm

mounting bracket so the Z-arm is secure.

8 While holding the adjustment tool in place,

slide the Z-arm off the tool. Ensure that the

bottom of the Z-arm lightly rubs against the

adjustment tool as it moves. Repeat steps 4

through 7 until this is true.

9 Store the Z-height adjustment tool.