Gilson 215 User Manual

Page 112

Advertising
background image

D-4

Appendix

D

Pipetting T

echniques

The following example describes an instance when

the above 8-step procedure is required. An operator

needs to dispense 50 µL of reagent into each of 20

test tubes. The tube-to-tube consistency of reagent

volume is critical. However, in order to save time,

the operator wishes to make a single 1 mL

aspiration from the reagent reservoir and then

make multiple dispenses. If the 8-step procedure

was not used, the last aliquot of the 20 dispenses

would be more diluted than the others. If the

procedure was used, the requirements of the

routine and the operator’s desire to save time could

both be met. (Recommended volumes in this

example are 15 µL air gaps and 100 µL excess

volume in Step 4.)