With liquid level, Appendix – Gilson 215 User Manual

Page 106

Advertising
background image

C-8

Appendix

C

GSIOC Commands

Q

I

Reads the XYZ travel range. Returns “axis=min/max”

where:

axis - X, Y or Z.

min - Lowest position in tenths of millimeters.

max - Highest position in tenths of millimeters.

The first time the command is sent, it returns the X range;

the second time, it returns the Y range; the third time, it

returns the Z range.

R

I

Reads front panel LED display. Returns “cccccccc” which

are the eight characters currently shown on the display. If

a character is non-printable, the hexadecimal ASCII code

for the character is returned.

S

I

Reads the command in the synchronization buffer. There

may be up to eight commands pending. Returns “|” if

buffer is empty.

Smm

B

Sends a synchronized buffered command (mm) that will

be executed when the instrument is quiescent. Sending a

command can overwrite unexecuted, existing commands.

If you send this command without indicating a parameter

(mm), the buffer is cleared.

Ta,b,c,d

B

Sets the Z-height (from the locator plate up, in tenths of

millimeters) for each probe, where:

a is the Z-height for probe A

b is the Z-height for probe B

c is the Z-height for probe C

d is the Z-height for probe D

If a Z-height is not specified, that probe will not move. For

example, “T500,,300”, sets the probe A Z-height to 500 and

probe C Z-height to 300.

T

I

Reads the last set Z height using the buffered T command

(from the locator plate up, in tenths of millimeters) for each

probe. Returns:

a is the Z-height for probe A

b is the Z-height for probe B

c is the Z-height for probe C

d is the Z-height for probe D

Advertising