Appendix – Gilson 215 User Manual

Page 105

Advertising
background image

C-7

Appendix

C

GSIOC Commands

N

I

Reads the liquid detector output. Returns “a,b,c,d” where:

a - leftmost probe (probe A)

b - probe B

c - probe C

d - probe D

For each status, you’ll see A for air or L for liquid.

Oabcd

B

Sets each probe speed value (in µm/s) where:

a = speed of probe A

b = speed of probe B

c = speed of probe C

d = speed of probe D

If the value is not specified, the probe speed value will

remain the same.

The default speed is 125000 µm/s.

O

I

Reads each probe speed value (in µm/s). Returns “a,b,c,d”

where:

a = speed of probe A

b = speed of probe B

c = speed of probe C

d = speed of probe D

P

I

Reads X and Y encoder positions in integral units of 0.1 mm.

q

I

Reads probe range values in the X direction for each probe

(a–d) in sequence where:

a = minimum X value/maximum X value

b = minimum X value/maximum X value

c = minimum X value/maximum X value

d = minimum X value/maximum X value

The range is displayed in integral units of 0.1 mm. The

values are based on the current probe pitch. Changing the

pitch will alter the ranges.