Front panel, Start button, Stop button – Gilson 215 User Manual

Page 42: Led display, Power indicator light, Front panel -2, Start button -2, Stop button -2, Led display -2, Power indicator light -2

Advertising
background image

3-2

Operation

3

Front Panel

Front Panel

The front panel of the liquid handler contains a

Start button, Stop button, LED display, and power

indicator light.

Start Button

The Start button can be used to home the XYZ-arm

when the liquid handler is first powered up or

when the motors for the XYZ-arm have been

relaxed. When pressed, the yellow LED lights.

Stop Button

The Stop button is a large touch-sensitive pad that

can be used to terminate a program and stop the

liquid handler from responding to any more

commands coming from the running program. This

button also relaxes the motors for the XYZ-arm so

that you can easily lift the probe and move the arm.

When pressed, the yellow LED light is turned off.

In a situation where an emergency stop is required,

pressing the Stop button immediately stops the

liquid handler. The Stop button is designed to be

sensitive enough that if you just brush it with your

hand it activates.

LED Display

The 8-character LED display shows the current

status of the liquid handler and any error codes as

they are encountered. Your program can also

contain instructions for showing 8-character

messages on the display when the program is run.

Refer to

Section 5, Troubleshooting

for a list of

current error codes and required actions.

Power Indicator Light

This indicator becomes lit when you turn on power

to the liquid handler using the power switch located

on the rear panel. Refer to the rear panel diagram

on page

2-16

if necessary.