Error messages, Error messages -2, Troubleshooting – Gilson 215 User Manual

Page 74

Advertising
background image

5-2

Troubleshooting

5

Error Messages

Error Description

Solution

15

NV-RAM checksum is invalid

• Send the buffered ~9 GSIOC command to

reset the NV-RAM and initialize to the

default value. Run 215SETUP.EXE or the

215 Setup Utility program

• Replace the main board

16

X scale factor is invalid

Contact the Gilson Customer Service Department.

17

Y scale factor is invalid

Contact the Gilson Customer Service Department.

20

X motor position error

Turn power off then on to the liquid handler.

21

Y motor position error

Turn power off then on to the liquid handler.

22

Z motor position error

Turn power off then on to the liquid handler.

24

X target less than minimum X

Send the immediate Q command using the

GSIOC Utility Program to read the travel range.

Correct the error in the program controlling the

liquid handler.

25

X target more than maximum X Send the immediate Q command using the

GSIOC Utility Program to read the travel range.

Correct the error in the program controlling the

liquid handler.

26

Y target less than minimum Y

Send the immediate Q command using the

GSIOC Utility Program to read the travel range.

Correct the error in the program controlling the

liquid handler.

27

Y target more than maximum Y Send the immediate Q command using the

GSIOC Utility Program to read the travel range.

Correct the error in the program controlling the

liquid handler.

28

Z target less than minimum Z

Send the immediate Q command using the

GSIOC Utility Program to read the travel range.

Correct the error in the program controlling the

liquid handler.

Error Messages

Advertising