Home phase, Home phase -13 – Gilson 215 User Manual

Page 53

Advertising
background image

3-13

Operation

3

Home Phase

Use the Home Phase tab to display the current X-

and Y-phase of the instrument.

Clicking Start causes the liquid handler to perform

the phase procedure. This procedure consists of the

liquid handler homing itself 10 times.

The liquid handler finds out where home is located

by “feeling” for the back and left walls of the unit.

The liquid handler expects to find these walls in the

same place each time. If it does not, you will get an

error. If this error was caused by an obstruction,

just clear the obstruction and try again. If the

problem does not clear or if a change is made to the

mechanics, you will probably need to repeat this

option to find the true home location.

Once the process completes, the spreadsheet

displays the values generated from each phase

procedure.

Configuring the Liquid Handler