Liquid Level Detector

Liquid Level Detector

Liquid level detector not detecting liquid level

• Ensure that the level sensing cables are plugged

in.

• Check sensitivity setting in the 215 Setup Utility

(see page

3-9

) and lower the percentage.

• Check if liquid is detectable. Liquid level detection

works only if there is electrical conductivity in

your liquid. Liquid level detecting will not work

with most non-polar liquids. For intermediate

polarity liquids and polar liquids, check the

sensitivity setting in the 215 Setup Utility.

• Call Gilson if this is caused by faulty circuitry.

Liquid level detector falsely detecting liquid
level

• Ensure that probes are installed correctly.

• Check sensitivity setting in the 215 Setup Utility

(see page

3-9

) and raise the percentage.

• Call Gilson if this is caused by faulty circuitry.